Lum v. City of Brewton12/2/2003
"16. Subject State. The operator must enter subject's state of residence using the appropriate two letter designation.
"17. Subject Driver License or Social Security Number. The operator must enter subject driver's license, social security number of Alabama I.D. number.
"18. Subject (Male/Female). The operator must designate the subject's gender using 'M' for male and 'F' for female.
"19. Subject Date Of Birth. The operator must enter the subject's date of birth using the format (MM/DD/YYYY) or (MM/DD/YY). An entry of 01/01/1900 will be used when the subject's date of birth cannot be obtained.
"(5) Report of Breath Alcohol Test Result.
"....
"(b) Draeger Alcotest 7110 MK III. For purposes of this regulation two (2) acceptable samples of breath shall constitute a breath alcohol test.
"1. The Draeger Alcotest 7110 MK III is programmed to require a second sample of breath be tested no less than two (2) minutes and not more than fifteen (15) minutes after the first sampling. The two samples must agree within +/-0.020 grams of ethanol per 210 liters of breath. If the two samples do not agree within +/-0.020 grams of ethanol per 210 liters of breath, the instrument is programmed to begin a second breath alcohol test procedure. The instrument will prompt the operator to have the subject provide two (2) additional samples of breath. Failure to provide two (2) acceptable samples of breath for the second breath alcohol test procedure will constitute a refusal of the whole test. The lowest of the breath alcohol results from the completed breath alcohol test procedure will be reported. A record of the breath alcohol results from the breath alcohol test will be stored in the electronic memory. The Alcotest 7110 MK III is programmed to generate a Certificate of Breath Alcohol Analysis.
"2. A complete breath alcohol test procedure shall incorporate two (2) calibration checks to verify the calibration of the Draeger Alcotest 7110 MK III. A 0.020 grams of ethanol per 210 liters of breath calibration check shall precede and a 0.080 grams of ethanol per 210 liters of breath calibration check shall follow each set of duplicate subject samples. The Draeger Alcotest 7110 MK III must produce a reading between 0.015 grams of ethanol per 210 liters of breath and 0.025 grams of ethanol per 210 liters of breath inclusive when a 0.020 grams of ethanol per 210 liters of breath calibration check vapor is introduced into the instrument to pass the calibration check. The Draeger Alcotest 7110 MK III must produce a reading between 0.076 grams of ethanol per 210 liters of breath and 0.084 grams of ethanol per 210 liters of breath inclusive when a 0.080 grams of ethanol per 210 liters of breath calibration check vapor is introduced into the instrument to pass the calibration check. The Alcotest 7110 MK III will discontinue the test sequence if the acceptable range for the 0.020 grams of ethanol per 210 liters of breath calibration check or the 0.080 grams of ethanol per 210 liters of breath calibration check is exceeded. 'Calibration Checks Before and After Test - OK' will be noted on the Certificate of Breath Analysis to indicate the criteria for an acceptable calibration check was met.
"3. An internal diagnostic check is performed by the Draeger Alcotest 7110 MK III one hundred twenty eight (128) times per second. The Draeger Alcotest 7110 MK III will store in memory a record of the diagnostics test per